Skip to main navigation Skip to search Skip to main content

Urinary neopterin quantification by reverse-phase high-performance liquid chromatography with ultraviolet detection

  • M.R. de Castro
  • , G.S. Di Marco
  • , D.Y. Arita
  • , L.C. Teixeira
  • , A.B. Pereira
  • , D.E. Casarini

    Research output: Contribution to journal › Article › Academic › peer-review

    Abstract

    Neopterin plays an important role in the malignant disease diagnostics. However, the methods employed in neopterin determination are generally difficult and/or time consuming. The aim of this work was to standardize a practical method to quantify neopterin using high-performance liquid chromatography-ultraviolet (HPLC-UV) and quantify it in patients with systemic lupus erythematosus (SLE). Urine was collected from healthy subjects (n=49), patients with inactive (n=15), active (n=28), and highly active SLE (n=6). The HPLC was performed using two coupled reverse-phase columns eluted with 150 mM sodium phosphate, pH 4.0, under a flow rate of 0.8 ml/min, with UV detector set at 353 nm and 100-fold diluted urines. The inter- and intra-assay studies presented an imprecision of 12.5% and 12.9% for quality controls of 3.94 and 1.1 mol/ml, respectively. Recovery from 79.5% to 82% was observed throughout the assay's linear range. Subjects with active (874.2±165.38 mol/mol creatinin) and highly active SLE (1753.8±453.9 mol/mol creatinin) showed three- and sixfold increased neopterin levels, respectively, compared to subjects with inactive SLE (314.3±121.3 mol/mol creatinin) and healthy subjects (294.6±178.6 mol/mol creatinin) (P
    Original languageEnglish
    Pages (from-to)275-283
    JournalJournal of Biochemical and Biophysical Methods
    Volume59
    Issue number3
    DOIs
    Publication statusPublished - 2004

    Keywords

    • systemic-lupus-erythematosus
    • disease-activity
    • serum neopterin
    • rejection
    • markers
    • assay

    Fingerprint

    Dive into the research topics of 'Urinary neopterin quantification by reverse-phase high-performance liquid chromatography with ultraviolet detection'. Together they form a unique fingerprint.

    Cite this